RCT in 80 ECMO patients needing CRRT: integrating the CRRT circuit into the ECMO circuit did not change median CRRT circuit lifespan versus a separate circuit (72 vs 71 h; p=0.52); 28-day mortality (32.5% vs 35%) and serious adverse events were similar.
RCTn = 80 patients
Tangjitaree et al. The comparison of circuit lifespan between integration and separation approach in extracorporeal membrane oxygenation patients requiring continuous renal replacement therapy support: a randomized controlled trial (E-CRRT Trial)